Single Vaporizer Anesthesia Machine

Single Vaporizer Anesthesia Machine

Features:
·Integrated breathing circuit with APL valve, excellent tightness.
·Replacing drive bellow is NOT required when operating adult/pediatric anesthesia.
Chat Now

Product Details

Features:

·Integrated breathing circuit with APL valve, excellent tightness.

·Replacing drive bellow is NOT required when operating adult/pediatric anesthesia.

·10.1 inch screen with no dead space.

·Applied with 1 high-precision vaporizers. (Selectatec).

·Real-time detection for loops, Paw, BPM, VT and compliance.

·Comprehensive monitoring capabilities include ventilation parameters monitoring, preset parameters, alarm indication and breathing waveform monitoring.

·ACGO available to connect with T/Bain circuit for open operation(Optional).

·Yoke system with up to 3 back up cylinders (A type with 11L size)(Optional).

Technical Specification:

Modes of Ventilation

VCV: Volume Controlled Ventilation

A/C: Assist/control Ventilation

SIMV: Synchronized Intermittent Mandatory Ventilation

Manual: Manual

PCV: Pressure Control Ventilation

STAND-BY: STAND-BY

Ventilator parameter

Working type: Electronic control, pneumatic driven

Tidal Volume range: 20-1500ml

Pressure range: 5-60cmH2O

Respiration Rate Range: 1-100bpm (1-40bpm under SIMV)

I: E Range: 4:1-1:8

Trigger Range: -20-20cmH2O(based on PEEP)

Minute volume range: >18L/Min

SIGH range: 0-5/100

System standard

Hypoxic guard system: N2O cut-off valve, O2 concentration >25%

Safety valve: <12.5kPa

Fresh gas compensation: 25-75L/Min

Flow meter: 4 tubes, cascade 0-15L/min O2, 

                     0.1-15L/min N2O, 0.1-15L/min air(Optional)

Volume of CO2 absorber: 2L

Vaporizer: 1/2(Optional) Selectable, maximum 2 mounts(Isoflurane/Enflurane/Sevoflurane/Halothane)

Monitoring

Display graphics: Waveforms of P-T, F-T, V-T, loops of P-V, V-F, F-P

Types: VT, MV, BPM, Paw, Lung compliance, Inspiratory Platform, FiO2

Alarm Limits

Audio/Visual: No tidal volume, MV, Paw, FiO2 limitation O2,

Air failure, AC power failure, Battery low

PEEP

Type: Integrated, Electronic control

Range: Air failure, AC power failure, Battery lo

Optional

1.AGSS

2.Suction Unit 

3.Touch screen


Inquiry