CE Approved Hospital Anesthesia Machine Equipment

CE Approved Hospital Anesthesia Machine Equipment

Applied for closed, semi-closed, semi-open inhalation anaesthesia, controlled or assisted ventilation, monitoring and displaying patients’ ventilation parameters, applied for child and adult
Chat Now

Product Details

Feature:

Applied for closed, semi-closed, semi-open inhalation anaesthesia, controlled or assisted ventilation, monitoring and displaying patients’ ventilation parameters, applied for child and adult

Compact anaesthesia ventilation system with the standard of ISO 8835-4. APL valve available. Good air tightness, easy to be operated, cleaned and disinfected.

(Anaesthesia agent transporting unit) meets the standard of ISO 8835-4 with concentration precisely controlled. Advanced temperature, flow and pressure compensation technologies.

Anaesthesia ventilator meets the standard of ISO 8835-5 with gas driven and electrical control, Colour TFT screen displaying sorts of monitoring parameters.


Ventilator Specification:

Modes of ventilation

IPPV: Intermittent Positive Pressure Ventilation

SIMV: Synchronized Intermittent Mandatory Ventilation

Manual: Manual

SIPPV: Synchronized Intermittent Positive Pressure Ventilation

PCV: Pressure Control ventilation

SIGH: SIGH

Ventilator parameter

Working type: Electronically controlled, pneumatic driven

Tidal Volume range: 20-1500ml

Pressure range: 5-80 mbar

Rispriation Rate Range: 1-99bpm (1-40bpm under SIMV)

I:E Range: 4:1-1:8

Ptrigger Range: -20-20cmH2O

Minute volume range: >18L/Min

 System Standard

Hypoxic guard system: N2O cut-off valve, O2 concentration<25%

Safety valve: <12.5kPa

Fresh gas compensation: 25-75L/min

Flow meter: Cascade, 0-10L/min

Volume of CO2 absorber: 1.5L

Vaporizer: 1 selectetac, maximum 2 mounts(Isoflurane/Enflurane/Sevoflurane/Halothane)

Alarm Limits

Audio/Visual: No tidal volume, MV, Paw, FiO2 limitation O2, Air failure, AC power failure, backup low

Pipeline supply

Gas configuration: O2, N2O, Air

Pipeline input range: 280-600kPa

Battery Power

Battery Type: Lead-acid 12V, 4Ah

Run-time: Typical 30 mins+, up to 2hrs

 Pre-setting functions

Language: ENGLISH

Calibration: Automatic

Applications

Patient range: Adult/Paediatrics

PEEP (Optional)

Type: Integrated, Electronic control

Range: OFF, 0-30cmH2O

Optional

ACGO: Auxiliary Common Gas Outlet

Vaporizer: Selectatec

Yoke system(Pin-index): Support 11L cylinder, O2/N20

AGSS: Anaesthetic Gas Scavenging Systems

Environmental Conditions

Operating Temp: 5℃~40℃

Storage Temp: -10℃~+40℃

Humidity: ≤80%

Atmosphere pressure: 960hpa~1040hpa


Inquiry