full automated chemistry analyzer 300T/hour
Model: CNME030405-2
Introduction
An automated analyser is a medical laboratory instrument designed to measure different chemicals and other characteristics in a number of biological samples quickly, with minimal human assistance. These measured properties of blood and other fluids may be useful in the diagnosis of disease.
Many methods of introducing samples into the analyser have been invented. This can involve placing test tubes of sample into racks, which can be moved along a track, or inserting tubes into circular carousels that rotate to make the sample available. Some analysers require samples to be transferred to sample cups. However, the effort to protect the health and safety of laboratory staff has prompted many manufacturers to develop analysers that feature closed tube sampling, preventing workers from direct exposure to samples.
Technical Specification:
Assay model | Random access with direct reading system | |
Method | End point, Kinetic, Fixed time, Immunoturbidimetry, 1-2 reagent, Multi standard, reagent/serum blank, etc | |
Assay Item | 40 assays | |
ISE mold (Na+ , K+ , Cl- /Li+) available on request | ||
Throughput | 1 Reagent | 300 tests/h(max) |
2 Reagent | 240 tests/h | |
With ISE | +120 tests/h | |
Test Menu | TBILI r-GT/GGT CHO Ca APOB-1/B CHE | |
D.BILI ALP/AKP TG CL APOB CREA | ||
TP UREA CK P CK-MB Mg | ||
ALB CREA LDH C ASO AFU | ||
Lactic Acid ALT/GPT UA α-HBDH CO2 FMN | ||
LP(a) AST/GOT GLU AMY LDL-C HDL-C | ||
CHE LA TBA ADA HS-CRP PA | ||
Fe Cu Zn C3 C4 LGG | ||
LGM LGA and drug, toxicity, etc | ||
Sample position | 60 | |
Including standard, control and stat; either serum cup or primary tube available | ||
Sample Volume | 1~100ul, 0.1ul/step | |
Reagent Position | 40 or 80(optional) | |
Reagent Volume | R1:1~400ul, 1ul/step, R2: 1~400ul, 1ul/step | |
Reagent/Sample probe | With liquid level sensor, collision sensor, Teflon coating and automatic washing station. | |
Reaction Time | 0~999 seconds | |
Cuvette | Washing High quality UV-transmitted plastic cuvette. Quartz glass cuvettes available on request. | |
Cuvette Washing | 8 channels automatic washing system for reaction cuvettes | |
Probe and Mixer Washing | Independent washing station for each sample/reagent probe and mixer | |
Alarm | Automatic alarm for malfunction (eg: collision, reagent/sample insufficiency, full waste container, etc) | |
Calibration | Liner, Non-liner, Multi standard, K factor | |
Wavelength | 340~810nm | |
Light Source | Halogen lamp | |
Absorbance Resolution | 0.0001A | |
Absorbance Accuracy | ±0.0003A(0~2.5A) | |
Repeatability (CV%) | ≤2% | |
Quality Control | Levy Jennings quality control program at 3 levels | |
Bar Code | Optional | |
Printer | Multiple report formats available | |
Test Control Function | New card control system available on request | |
Language | English, French, Spanish, German, Russian, etc | |
Power Supply | 220V/50Hz or 110V/60Hz (optional) | |
Dimension | 85*72*53(cm) | |
Weight | 90kg |